Welcome to the official website of Hangzhou Sicheng Environmental Protection Technology Co., Ltd!

Focusing on environmental protection and industrial online monitoring products

400-666-5501

CN/EN

Home >> Product >>Online monitoring of gas pollution sources >> GCOM-520 volatile organic compound online analyzer (gas chromatograph)
Product Detail

GCOM-520 volatile organic compound online analyzer (gas chromatograph)

Collect
  • trade description

Product Introduction:

The GCOM-3220 volatile organic compound online analyzer (gas chromatograph) uses a dedicated chromatographic column combination, gas path blowback technology, and dual hydrogen flame ionization detector (FID) technology to achieve simultaneous detection of non methane total hydrocarbons and benzene derivatives. It is suitable for high temperature, high humidity, and high dust conditions, and the data can be uploaded to the corresponding monitoring platform according to requirements.

Product Features:

The analyzer adopts GC+dual FID detection technology to achieve simultaneous monitoring of non methane total hydrocarbons and benzene derivatives;

The entire process of sample collection, transmission, and monitoring is accompanied by heat, ensuring accurate and reliable measurement data;

Using MFC electronic flow control technology to control gas flow, with high control accuracy and good stability;

The dual detector configuration, dual column box structure design, and support for multi-point continuous detection have improved the application range;

The intelligent FID detector has a flameout and self ignition alarm function, which effectively controls the safety of the hydrogen flame gas path;

Automatic cycle operation upon startup, automatic fault diagnosis, regular automatic calibration, truly achieving online analysis;

Support remote control of the analyzer via TCP/IP network, and the analyzer can also be directly operated on-site through the touch LCD screen.

Product Application:

Petrochemical, organic chemical, surface coating, printing and packaging, coatings and packaging, electronics industry, furniture manufacturing, shipbuilding industry, automobile manufacturing and other fields.

Product scale:

Measurement method: GC dual FID detection technologyAnalysis cycle: Non methane total hydrocarbons ≤ 1 minute, benzene derivatives ≤ 3 minutes (xylene not separated) or ≤ 8 minutes (xylene separated)

Measurement range: Total hydrocarbons, methane, non methane total hydrocarbons, benzene, toluene, xylene and other benzene derivatives Range: Non methane total hydrocarbons: 0.05 μ mol/mol~1000 μ mol/mol, range can be customized

Benzene derivatives: 0~10 μ mol/mol, adjustable range

Sample gas input flow rate:100mL/min
Air input parameters: hydrocarbon free air, hydrocarbon content ≤ 0.3mg/m3, (0.30 ± 0.01) MPa
Detection limit: Non methane total hydrocarbons: ≤ 0.05 μ mol/mol; Benzene derivatives: ≤ 0.1 μ mol/mol Repeatability: RSD ≤ 3.0%Hydrogen input parameters: H2, ≥ 99.999%, (0.30 ± 0.01) MPa
Carrier gas input parameters: N2, ≥ 99.999%, (0.30 ± 0.01) MPa
Communication output: 10/100M Ethernet port, RS232Auxiliary gas input parameters: oil-free compressed air, 0.45MPa~0.6MPa
Communication protocol: Modbus RTU, HJ 212 or local protocolDimensions: 483mm (W) × 221mm (H) × 600mm (D)
Working temperature: 5 ℃~40 ℃Power supply: 220 ± 10VAC, 50 ± 1Hz, 900W (maximum)


×