Welcome to the official website of Hangzhou Sicheng Environmental Protection Technology Co., Ltd!

Focusing on environmental protection and industrial online monitoring products

400-666-5501

CN/EN

Home >> Product >>Online monitoring of water pollution sources >> BLA-TPN-500 Total Nitrogen Water Quality Online Analyzer
Product Detail

BLA-TPN-500 Total Nitrogen Water Quality Online Analyzer

Collect
  • trade description
  • brief introduction

The BLA-TPN-500 total nitrogen water quality online automatic analyzer is a new type of fully automatic online analyzer used to measure total nitrogen in sewage. This product is based on our patented magnetic conductivity measurement platform, combined with self-developed xenon lamp/grating spectrometer technology, digestion colorimetric integration technology, etc., to achieve high-precision, low detection limit, high stability, and low maintenance total nitrogen fully automatic online monitoring. The working principle of this product complies with the technical requirements of HJ 636-2012 "Determination of total nitrogen in water quality - Alkaline potassium persulfate digestion UV spectrophotometry", fully meeting the applications in pollution sources, industrial production processes, municipal sewage and other fields. Certified by CCEP.

  • characteristic

       Magnetic conductivity measurement technology is not affected by interference such as high chromaticity, suspended solids, and bubbles;

       High reliability and strong resistance to turbidity interference;

       High detection accuracy, low quantification limit (0.1mg/L), and small long-term drift of the instrument;

       Equipped with one click self-test function and self diagnostic function;

       Single channel highly integrated valve island, capable of single valve maintenance, disassembly, and cleaning;

       Scalable 24/7 networking function, monitoring instrument operation status anytime, anywhere;

       Equipped with alarm function, quality control function, and reverse control function.

  • application

Widely used in environmental pollution source monitoring, industrial process water monitoring, municipal sewage treatment, surface water monitoring, etc.

  • specifications

       Measurement method: Alkaline potassium persulfate digestion UV spectrophotometry

       Photometric method

       Range: 0.1~5/25/50/200 mg/L (expandable)

       Indication error: ± 10%

       Repeatability: ≤ 5%

       Zero drift: ± 5% F.S

       Range drift: ≤ 5% F.S

       Quantitative lower limit: 0.1mg/L

       Reagent dosage: Standard reagents for 400 samples

       Measurement interval: whole point measurement, interval measurement, or through serial port

       Start measurement

       Maintenance workload:<2 hours/month

       Environmental temperature: 5 ℃~45 ℃

       Communication interface: RS485/RS232/USB interface (optional)

       Standard configuration includes 1 4-20mA input and output (expandable to 2 channels)

       Power consumption: 200W (220 ± 10% VAC 50Hz), excluding water collection pump

       Note: The testing method refers to the latest environmental certification standards

        

×