BLA-TP-500 Total Phosphorus Water Quality Online Analyzer![]()
Collect
The BLA-TP-500 total phosphorus water quality online automatic analyzer is a new type of fully automatic online analyzer used to measure total phosphorus in sewage. This product is based on our patented magnetic conductivity measurement platform, combined with constant temperature optical fiber, digestion colorimetric integration technology, etc., to achieve high-precision, low detection limit, high stability, and low maintenance total phosphorus fully automatic online monitoring. The working principle of this product complies with the technical requirements of GB 11893-1989 "Determination of total phosphorus in water quality - Ammonium molybdate spectrophotometric method", fully meeting the applications in pollution sources, industrial production processes, municipal sewage and other fields. Certified by CCEP.
Magnetic conductivity measurement technology is not affected by interference such as high chromaticity, suspended solids, and bubbles; High reliability and strong anti-interference ability; High detection accuracy, low quantification limit (0.05mg/L), and small long-term drift of the instrument; Equipped with one click self-test function and self diagnostic function; Single channel highly integrated valve island, capable of single valve maintenance, disassembly, and cleaning; Scalable 24/7 networking function, monitoring instrument operation status anytime, anywhere; Equipped with alarm function, quality control function, and reverse control function.
Widely used in environmental pollution source monitoring, industrial process water monitoring, municipal sewage treatment, surface water monitoring, etc.
Measurement method: Ammonium molybdate spectrophotometric method Range: 0.05~2/10/50mg/L (expandable) Indication error: ± 10% Repeatability: ≤ 5% Zero drift: ± 5% F.S Range drift: ≤ 5% F.S Quantitative lower limit: 0.05mg/L Reagent dosage: Standard reagents for 400 samples Measurement interval: whole point measurement, interval measurement, Measurement can also be initiated through serial port Maintenance workload:<2 hours/month Environmental temperature: 5 ℃~45 ℃ Communication interface: RS485/RS232/USB interface (optional) Standard configuration includes 1 4-20mA input and output (expandable to 2 channels) Power consumption: 200W (220 ± 10% VAC 50Hz), excluding water collection pump Note: The testing method refers to the latest environmental certification standards |